| Model Number |
E2750 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/30/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date of initial report: 12/15/2016.To date, the incident sample has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the activation button keeps sticking on the handset.There was no injury to the patient.The issue was solved by using a new device.
|
| |
|
Additional Manufacturer Narrative
|
|
Covidien reference#: (b)(4).Date of initial report: 12/15/2016.Date of follow-up report: 01/20/2017.Per engineering investigation, the ¿latch on¿ condition was caused by interference between the rocker component and the rocker window in which it moves.The width of the rocker was found to be larger than the width of the rocker window.The width of the rocker window is formed by the ultrasonic welding of the two body housing components.Corrective action is being issued.
|
| |
|