MAUDE Adverse Event Report: CYBERONICS, INC. LEAD MODEL 302

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problems High impedance (1291); Failure to Interrogate (1332); Mechanical Problem (1384); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2016
Event Type  malfunction  
Event Description
During a generator replacement surgery, the old generator was unable to be interrogated.When the lead was connected to the new generator, high impedance was observed.The company representative performed 3 system diagnostics in a row.Each time, the output current was low, current delivered was 0.25ma, and lead impedance was high >=10000 ohms.The old lead pin was fully inserted into the m 103 prior to system diagnostics.The surgeon looked under magnification to confirm, then disconnected lead, inspected lead and insertion point and confirmed again.High impedance was still present after confirming complete insertion.A full revision surgery was performed as a result.With the new lead connected to the new generator, impedance was within normal limits.The explanted lead has not been received to date.
 
Event Description
The generator and lead were received.Analysis is underway but has not been completed yet.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The reported ¿end of service and low battery¿ allegations on the generator were confirmed in the lab; an open can measurement of the battery voltage determined that the battery was depleted.The end of service condition was the result of normal, expected battery depletion based on the battery life calculation, the electrical test results and the bench evaluation.Analysis of the generator in the lab concluded that no abnormal performance or any other type of adverse condition was found.An analysis was performed on the returned lead portions and the reported allegations of lead fracture and high impedance were confirmed.A portion of the outer silicone tubing and a portion of the connector ring quadfilar coil were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned lead, the lead coils appeared to be broken in several areas.Scanning electron microscopy performed on the several coil breaks identified the areas as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage, some with pitting and some without pitting, evidence of a stress induced fracture (rotational forces) which most likely completed the fracture on two of the broken coil strands.The area on one of the two remaining broken coil strands was identified as having extensive pitting which prevented identification of the coil fracture type.The area on the last broken coil strand was identified as having evidence of being worn to the point of fracture which prevented identification of the coil fracture type.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.During the visual analysis of the returned 15mm portion the (-) green quadfilar coil appeared to be broken approximately 3mm past the end of the cut / torn inner silicone tubing.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.The slice mark found on the outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids found inside the outer and inner silicone tubing.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed on the deposit observed on the inner silicone tubing and identified the deposit as containing silicon, phosphorus, sodium, magnesium and calcium.With the exception of the observed discontinuities the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6179592
MDR Text Key62865781
Report Number1644487-2016-02885
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2006
Device Model Number302-20
Device Lot Number010763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2016
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received01/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-