Device Problems
Moisture Damage (1405); Structural Problem (2506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging that the battery compartment was found to be cracked and moisture was present.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
The device has been returned and evaluated by product analysis on 12/30/2017 with the following findings: during visual inspection of the pump, it was observed that the battery compartment had two cracks and there was evidence of moisture in the compartment, on the motor flex connector.A leak test was performed and failed due to the battery compartment leak.The pump was opened and there was no further moisture damage found internally.
|
|
Manufacturer Narrative
|
The correct date of evaluation by product analysis is 12/30/2016.
|
|