Medtronic received information that during treatment of an aneurysm located in the right internal carotid artery (ica), a dissection, resulting in the placement of a stent, occurred within the right ica.Per the event report the dissection was likely a result of the number of passages made to deliver the device and the time of the procedure as the patient had severe vessel tortuosity.
|
As received, the pipeline flex delivery system (ped-425-20) was returned stuck inside the distal segment of the microcatheter.The distal hypotube appeared to be stretched.The distal end of the pipeline braid was observed slightly frayed.The pushwire was found bent and kinked.The catheter body was found to have accordion damage.No other anomalies were observed.Based on the reported information and device analysis we are unable to definitively determine the cause for the reported experienced.However, it is possible that the ¿severe vessel tortuosity¿ may have contributed to ¿resistance¿ during delivery; subsequently causing the pipeline flex delivery system and the catheter to become damaged and stuck.However, the cause for resistance could not be determined.Additionally, the damages seen on the catheter body (accordioning), proximal wire (kinking/bending), pipeline braid (fraying) and hypotube (stretching); suggest excessive force was used.Furthermore, it was reported that the dissection of the right ica was likely a result of the number of attempts made to deliver the pipeline.We are unable to definitively determine the cause for the dissection.However dissection is listed as a potential complication in the pipeline flex ifu.Per our instructions for use (ifu), the user should: ¿discontinue delivery of the device if high force or excessive friction is encountered during delivery.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intralumenal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Movement of the micro catheter against resistance may result in damage to the micro catheter, or the vessel.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ all products are 100% inspected for damage and irregularities during manufacture.
|