Both of the faulty ups units were returned by the customer to merge healthcare on 02dec2016 for evaluation.The results showed that the batteries were defective and subsequently replaced.Upon replacing the batteries, the units passed all qc testing and were then sent to service stock.Device labeling, hemo v9.40 user manual, addresses the potential for such an occurrence with statements such as, "if low voltage is detected, the pdm goes to battery power and an audible alarm sounds.".
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the ups (uninterrupted power supply) was buzzing and caused the hemo application to shut down during two (2) procedures, each in different labs at the site.Information obtained from the customer revealed that upon failure of the ups units, the user immediately plugged the hemo system into a wall outlet and then waited for the system to reboot.The delay was ~10-15 minutes.This resulted in a loss of patient monitoring during the reported delay.If parts of the hemo system become energized, there is a potential for direct harm to the patient and/or user including electrical shock or burns.However, the customer confirmed that there were no injuries to the patients or staff.With merge hemo not capturing physiological data, there is a potential for delay of treatment that could cause harm to the patient.However, the customer reported that the procedures were completed successfully once the hemo system was rebooted.(b)(4).
|