| Catalog Number |
K15B |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/21/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
X-ray review: post op kyphoplasty images for l4/5 kyphoplasty provided.Balloon markers are present in l5 owing to balloon rupture.No information is provided about expansion pressure.Balloon can be ruptured if over inflated past instructions for use(ifu) recommended pressure or if pulled back against "pac" needle.Root cause: surgical technique.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2016 the patient underwent kyphoplasty at l-5.Intra-op, the balloon popped during inflation.While pulling the balloon out, some part of it remained inside the vertebral body.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis:the reported complaint of a balloon rupture was confirmed during product evaluation.A radial tear was observed near the proximal peak of the balloon.This observation is consistent with rupture due to contact with bone splinters during instrument usage.Due to the ¿umbrella¿ effect of the radially cut ibt, the ibt was unable to be collapsed and removed from the stylus.The above observations are consistent with intraoperative instrument damage.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|