| Model Number |
M0061751970 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A visual analysis of the returned percuflex¿ plus ureteral stent found that the stent was broken at the middle section.It is possible that the way in which the device was handled and manipulated during unpacking or preparation may have contributed to the failure encountered (stent broken).During manufacturing the units are inspected in order to avoid the failures found.Most likely, the damages found are consistent with one caused when catheter is being pulled; this may cause damage, deformities or device break.Therefore, the most probable root cause assigned to the complaint is ¿handling damage.¿ the device history record (dhr) review found that the device met all manufacturing specifications.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a percuflex¿ plus ureteral stent was used during a procedure performed on an unknown date.According to the complainant, during the procedure and outside the patient, the tip of the stent broke.No harm was noted to the patient.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the stent shaft was found broken.
|
| |
|