| Model Number |
H7493895940200 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
11/25/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and vessel perforation occurred.The target lesion was located in the posterior descending artery (pda).A 2.00mm x 40mm apex¿ balloon catheter was advanced for dilatation.However, during the first inflation at 12 atmospheres, the balloon ruptured.A perforation on the artery was then noted and a stent was deployed to cover the perforation.No patient complications were reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned device consisted of a apex balloon catheter.The balloon was loosely folded, with blood and in the wire lumen, hub and balloon.The tip, balloon, hypotube and distal shaft were microscopically and tactile inspected.Inspection revealed a small pinhole in the balloon material at the distal edge of the distal markerband with numerous kinks in the hypotube.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and vessel perforation occurred.The target lesion was located in the posterior descending artery (pda).A 2.00mm x 40mm apex¿ balloon catheter was advanced for dilatation.However, during the first inflation at 12 atmospheres, the balloon ruptured.A perforation on the artery was then noted and a stent was deployed to cover the perforation.No patient complications were reported and the patient's status was fine.
|
| |
|
Search Alerts/Recalls
|