As reported per mdr (b)(4): "the patient originally underwent a surgery for l3 4 l4 5 decompressive laminectomy for lumbar spinal stenosis with posterior stabilization using the coflex interspinous process clamp on (b)(6) 2016.No complications were reported during the original procedure.The patient did return for hospitalization on (b)(6) 2016 for development of drainage from the incision site that did not subside.On (b)(6) 2016 the patient again returned to the hospital for a surgical procedure related to loose orthopedic implant and removal of the posterior spinal instrumentation.The neurosurgeon documented that the patient recently was seen for with complaints of increasing lower back pain associated with movement.The documentation revealed that it was somewhat more on the left side than the right and x-rays were obtained that showed the coflex device had migrated posteriorly and was no longer in the interspinous space.Removal was done in hopes of helping the patient with his residual back pain.The neurosurgeon documented that flexion extension x-rays failed to show any obvious instability.Post-operative mri scan showed successful decompression of the spinal canal.No further information was available and no specific documentation was found that the devices had specific concerns.".
|