MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE PROX EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE PROX EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1076250-08
Device Problems Difficult To Position (1467); Difficult to Remove (1528); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that during a coronary procedure, resistance was met with the guide catheter during advancing and removal of the device.Once removed, the stent struts were noted to be flared.A different xience prox was used successfully in the procedure.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: visual, dimensional and functional inspections were performed on the returned device.The difficult to position with the guiding catheter and stent damage were confirmed.The difficulty removing the stent delivery system could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.The xience prox is currently not commercially available in the us; however, it is similar to a device sold in the us.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE PROX EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6187751
MDR Text Key63190667
Report Number2024168-2016-09022
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2019
Device Catalogue Number1076250-08
Device Lot Number6080441
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2016
Initial Date FDA Received12/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-