• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012448-20
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/24/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.The other 3 nc trek devices referenced in b5 are filed under separate medwatch reports.
 
Event or Problem Description
It was reported that prior to use, during device preparation, the protective sheath of the nc trek balloon met resistance during removal.Additionally, the device was not submerged in sterile heparinized normal saline to activate the coating; however, the device was successfully used in the patient without reported issue.This occurred with 4 separate devices.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6187880
Report Number2024168-2016-09031
Device Sequence Number716895
Product Code LOX
UDI-Device Identifier08717648151835
UDI-Public(01)08717648151835(17)190630(10)60716G1
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number1012448-20
Device Lot Number60716G1
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/25/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/19/2016
Supplement Report FDA Received Date04/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-