No patient information provided as no patient was involved in this concern.A part shipment was placed for a position sensor unit (psu), sensor control unit (scu) and two connection cables.A medtronic representative went to the site to test the equipment.While performing a navigation system check-out, the medtronic representative, confirmed that the navigation system psu would not power on.There was no optical communication so the optical procedure could be performed.The medtronic representative deduced the psu required replacement.After psu replacement, the medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.Return requested for suspect positioning sensor unit (psu).Manufacture date apr 2012.The returned psu powered up without the amber fault light on.The psu functioned normally for several minutes, and then the fault light came on.A check of the event log revealed a history of illuminator current failure.The illuminator current has been moving in and out of range since (b)(6) 2013.The psu failed an aak test at.50 mm with a passing threshold of.35 mm.The laser battery is also very low.The hardware investigation was completed and revealed that this issue was related to a hardware issue and was documented in a medtronic hardware anomaly tracking database.The sensor control unit (scu) and two connection cables were returned unused.
|