BOSTON SCIENTIFIC - MAPLE GROVE APEX¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493895915200
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/07/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years old or older.(b)(4).Device evaluated by mfr.: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary artery.A 2.00mm x 15mm apex¿ balloon catheter was advanced for dilatation.However, upon introduction, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a apex balloon catheter.The balloon was tightly folded.There was contrast in the inflation lumen.There was blood in between balloon folds.The outer shaft, inner shaft, balloon and tip were microscopically examined.There were numerous hypotube kinks.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary artery.A 2.00mm x 15mm apex¿ balloon catheter was advanced for dilatation.However, upon introduction, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key6190704
Report Number2134265-2016-12086
Device Sequence Number877007
Product Code LOX
UDI-Device Identifier08714729475590
UDI-Public(01)08714729475590(17)20190430(10)19121482
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model NumberH7493895915200
Device Catalogue Number38959-1520
Device Lot Number19121482
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 12/08/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/20/2016
Supplement Report FDA Received Date01/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-