| Model Number |
25-28-67 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/24/2016
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Sorin group (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The gas blender has been returned to sorin group (b)(4) for investigation.A follow-up will be sent when the investigation is complete.
|
| |
|
Event or Problem Description
|
|
Sorin group (b)(4) received a report that the s5 gas blender system displayed an error code during installation.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The model number provided in the initial report (25-40-45) was incorrect.The correct model number is 25-28-67.
|
| |
|
Additional Manufacturer Narrative
|
|
Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The device was returned to livanova for further investigation.The described error could be reproduced.It was discovered that there was a loose plug connection at the co2 valve.The parts were replaced.The unit was disinfected and cleaned.Preventative maintenance was performed and the unit function properly without any further issue.
|
| |
|