• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Codes Device Operates Differently Than Expected (2913); Activation Failure (3270)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/22/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter stated the surgeon inserted and removed a 13.2mm micl13.2 implantable collamer lens, -16.0 diopter, several times during the same surgery due to the lens would not unfold in the patient's eye.The lens eventually was torn and the incision was enlarged to remove the lens.The backup lens was implanted and sutures were not required.There were no additional complications and the patient is doing well.The reporter indicated the surgeon had problems folding the lens during the loading process and the event was due to surgeon/technique error.There were no issues with the lens or injection system.
 
Additional Manufacturer Narrative
Visual inspection of the returned product found the lens optic torn and a piece of one haptic torn off and missing.The lens was returned in liquid.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key6191981
Report Number2023826-2016-01784
Device Sequence Number415872
Product Code MTA
UDI-Device Identifier00841542103411
UDI-Public(01)00841542103411(17)170630
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date06/30/2017
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/22/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/20/2016
Supplement Report FDA Received Date02/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45 FP - LOT NUMBER UNK; FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
-
-