| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Device Operates Differently Than Expected (2913); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/22/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated the surgeon inserted and removed a 13.2mm micl13.2 implantable collamer lens, -16.0 diopter, several times during the same surgery due to the lens would not unfold in the patient's eye.The lens eventually was torn and the incision was enlarged to remove the lens.The backup lens was implanted and sutures were not required.There were no additional complications and the patient is doing well.The reporter indicated the surgeon had problems folding the lens during the loading process and the event was due to surgeon/technique error.There were no issues with the lens or injection system.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned product found the lens optic torn and a piece of one haptic torn off and missing.The lens was returned in liquid.(b)(4).
|
| |
|