Model Number U128 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285); Ambient Noise Problem (2877)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/28/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
|
Event Description
|
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) and another manufacturer's right atrial (ra) lead exhibited low out of range pacing impedance measurements less than 200 ohms.Additionally, the crt-p and another manufacturer's right ventricular (rv) lead exhibited oversensing of myopotential noise that resulted in less than 2 seconds of pacing inhibition.Boston scientific technical services (ts) discussed troubleshooting and programming options.No adverse patient effects were reported.This product remains implanted and in service.
|
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
Additional information was received that the patient was seen in clinic for evaluation of the ra lead.Patient isometrics were performed and noted stable pacing impedance measurements, however, noise was able to be reproduced in bipolar and unipolar configurations.The ra lead was left programmed unipolar and sensitivity was decreased.The root cause was not determined and the physician plans to continue monitoring.
|
|