• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2016
Event Type  malfunction  
Manufacturer Narrative
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) and another manufacturer's right atrial (ra) lead exhibited low out of range pacing impedance measurements less than 200 ohms.Additionally, the crt-p and another manufacturer's right ventricular (rv) lead exhibited oversensing of myopotential noise that resulted in less than 2 seconds of pacing inhibition.Boston scientific technical services (ts) discussed troubleshooting and programming options.No adverse patient effects were reported.This product remains implanted and in service.
 
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Additional information was received that the patient was seen in clinic for evaluation of the ra lead.Patient isometrics were performed and noted stable pacing impedance measurements, however, noise was able to be reproduced in bipolar and unipolar configurations.The ra lead was left programmed unipolar and sensitivity was decreased.The root cause was not determined and the physician plans to continue monitoring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6194546
MDR Text Key62978113
Report Number2124215-2016-18679
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/27/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4674; 5024M; 5524M; SEDR01; U128
Patient Age32 YR
-
-