| Model Number |
BL5425V |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/09/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sterilization and lot history records were reviewed and found to be acceptable.Report 1 of 3 see 1920664-2016-00479, and 00480.
|
| |
|
Event or Problem Description
|
|
The cutter stops working in mid stream.The surgeon has to lift up his foot from the pedal and then press it down again for the cutter to work.Additional information: the cutter is at 5000cpm when this occurs.This has been happening intermittently over time, no date of event.There has been no injury to patients and no need for medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation completed.Received sealed bl5425v pack from lot v7875.The expiration date on the label is 2018-02-07.The returned sample was opened for inspection and tested.No defects were found.A functional test was performed using a stellaris pc system.The cutter had good clean cuts and had continuous cutting.The cutter did not stop during functional testing.Report 1 of 3 see 1920664-2016-00479, 00480.
|
| |
|