| Catalog Number |
NF1330 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/12/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).One unused set was received for evaluation.The set was tested for leakage per the instructions for use.There was no leakage noted.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.The returned set met requirements according to specification, and the reported failure could not be reproduced.If additional pertinent information becomes available a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported by user facility: access port (for needleless entry) leaking blood when placed on iv angiocath and blood being drawn with a luer-lock syringe.No prior access from access port.
|
| |
|