MAUDE Adverse Event Report: STRYKER SPINE-FRANCE OASYS POLYAXIAL SCREW NON BIASED DIA 3.5 X 12; SPINAL INTERLAMINAL FIXATION ORTHOSIS.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-FRANCE OASYS POLYAXIAL SCREW NON BIASED DIA 3.5 X 12; SPINAL INTERLAMINAL FIXATION ORTHOSIS. Back to Search Results
Catalog Number 48558312
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/29/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a screw on the left at c5 had to be removed, after removing blocker and getting rod out of the way, the surgeon was unable to access the screw head due to it being cold fused.Rocked screw back and forth too much, then used an alternative instrument and pulled screw out.This was not ideal and surgeon requested i send screw in to be analyzed.
 
Additional Manufacturer Narrative
Method: visual inspection; device history review; complaint history review; risk assessment; results: the screw was examined and found to have a tulip head that was jammed into the screw shank.Manufacturing files were reviewed and no anomalies were found.Conclusion: the probable cause is patient factors due to the screw being implanted and a constant force applied to the tulip head jamming it into the shank.
 
Event or Problem Description
It was reported that a screw on the left at c5 had to be removed, after removing blocker and getting rod out of the way, the surgeon was unable to access the screw head due to it being cold fused.Rocked screw back and forth too much, then used an alternative instrument and pulled screw out.This was not ideal and surgeon requested i send screw in to be analyzed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OASYS POLYAXIAL SCREW NON BIASED DIA 3.5 X 12
Common Device Name
SPINAL INTERLAMINAL FIXATION ORTHOSIS.
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
MDR Report Key6196350
Report Number0009617544-2016-00500
Device Sequence Number509699
Product Code KWP
UDI-Device Identifier04546540596314
UDI-Public(01)04546540596314
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K072568
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48558312
Device Lot Number161490
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/29/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/21/2016
Supplement Report FDA Received Date02/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-