| Catalog Number |
7570855 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/15/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.
|
| |
|
Event or Problem Description
|
|
It was reported patient underwent an unspecified spinal procedure in which product was found damaged.Physician had to replace it with new one to complete the surgery successfully.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis :macroscopic and optical examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|