• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) TACTILE AWL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) TACTILE AWL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734435
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information not provided due to (b)(6) patient privacy regulations.No parts have been received by manufacturer for analysis.
 
Event or Problem Description
A medtronic representative reported that while in a spinal fusion the tactile awl became bent.The surgeon had the tactile awl placed in the pedicle, they tried to change directions and then after removing discovered the damage, there was no reported delay to the procedure due to this issue.There was no impact on patient outcome.
 
Additional Manufacturer Narrative
The suspect awl was returned to the manufacturer for evaluation.Testing found that the tip of the instrument was bent.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTILE AWL
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key6198119
Report Number1723170-2016-05899
Device Sequence Number18845232
Product Code OLO
UDI-Device Identifier00613994759610
UDI-Public00613994759610
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number9734435
Device Lot Number150915
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/29/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/22/2016
Supplement Report FDA Received Date06/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-