It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 2.75x24mm promus element ¿ drug-eluting stent was advanced to treat the lesion.However, during the procedure, it was noted that the stent struts were deformed.No patient complications were reported and the patient's status was stable.
|
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.A visual examination of the crimped stent found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal marker bands.The crimped stent outer diameter (od) is within the specification.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.The bumper tip of the device showed slight signs of damage to the distal edge of the tip.This type of damage is consistent with resistance being encountered on advancement of the stent delivery catheter to the lesion site.A visual and tactile examination found several hypotube kinks across the length of the shaft.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the mid-shaft section found several mid-shaft kinks across the length of the polymer shaft.This type of damage is likely to have been caused by excessive tensile force being applied to the delivery system.The outer extrusion, inner lumen and bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 2.75 x 24 mm promus element drug-eluting stent was advanced to treat the lesion.However, during the procedure, it was noted that the stent struts were deformed.No patient complications were reported and the patient's status was stable.
|