MEDTRONIC NEUROSURGERY DELTA SHUNT KIT, REGULAR, PERFORMANCE LEVEL 1.5; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
|
Back to Search Results |
|
| Catalog Number |
25132-5 |
| Medical Device Problem Codes |
Break (1069); Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Code |
Therapeutic Response, Decreased (2271)
|
| Date of Event |
11/21/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for siphon, reflux, and leak testing.However, it did not meet the requirements for pressure flow and pre-implantation testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ approximately 9.5 cm of the ventricular catheter was returned.It met the requirements for patency and leak testing.The peritoneal catheter was returned in two pieces of lengths 9 cm and 88 cm respectively.Proteinaceous debris was observed within the exterior of the catheter and the tear site appears to be jagged.It is unknown how or when this damage occurred.The returned segments were patent and passed leak testing when tested individually.The catheter met all of the requirements for the tensile strength and ultimate elongation tests.The instructions for use that accompany the device cautions that low tear strength is a characteristic of most unreinforced silicone elastomer materials, and that care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks, or tears.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.All catheters are 100% inspected at the time of manufacture.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic neurosurgery that the device was implanted in (b)(6) 2015.According to the report, it was found that the flow through the device was poor and when checked, it was determined the peritoneal catheter was broken.Reportedly, the system was removed including the broken parts of the catheter.The system was replaced and the patient was well.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|