MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012450-15
Device Problems Inflation Problem (1310); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4). During processing of this complaint, attempts were made to obtain complete event, patient and device information. The device was not returned for evaluation. The investigation was unable to determine a conclusive cause for the reported complaints. A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided. Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue. Based on the information reviewed there is no indication of a product quality issue with respects to the design, manufacturing, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a lesion in the circumflex artery. A 3. 25x15mm nc trek rx balloon dilatation catheter (bdc) stylet was difficult to remove. The bdc was advanced toward the target lesion, an attempt was made to inflate the balloon and the balloon completely failed to inflate. The bdc was removed and a same size nc trek was used to continue the procedure. There was no adverse patient effect and no clinically significant delay in the procedure. No additional information was provided.
 
Manufacturer Narrative
(b)(4). On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues. Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis. The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints. To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda [medwatch # 2024168-2017-02310]. Corrective action has been implemented per site operating procedures. The product will continue to be trended. The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC TREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6200324
MDR Text Key63489593
Report Number2024168-2016-09163
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/25/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number1012450-15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-