Catalog Number 1012450-15 |
Device Problems
Inflation Problem (1310); Difficult to Remove (1528)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/29/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.The investigation was unable to determine a conclusive cause for the reported complaints.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.Based on the information reviewed there is no indication of a product quality issue with respects to the design, manufacturing, or labeling of the device.
|
|
Event Description
|
It was reported that the procedure was to treat a lesion in the circumflex artery.A 3.25x15mm nc trek rx balloon dilatation catheter (bdc) stylet was difficult to remove.The bdc was advanced toward the target lesion, an attempt was made to inflate the balloon and the balloon completely failed to inflate.The bdc was removed and a same size nc trek was used to continue the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
|
|
Search Alerts/Recalls
|