Catalog Number 498-01-048
Health Effect - Clinical Code Pain (1994)
Date of Event 11/29/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient presenting with pain.The surgeon felt a liner swap would relieve that.
 
Additional Manufacturer Narrative
The reason for this revision surgery was pain.The in-vivo length of patient service for the implant was 11.3 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The root cause for the pain was not reported.The complaint states the surgeon believed a poly exchange would remedy the patient's condition.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KERAMOS, CERAMIC/CERAMIC HIP
Common Device Name
CUP, AL-AL FLARED RIM FOR CERAMIC INSERT 48MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6200862
Report Number1644408-2016-01030
Device Sequence Number323339
Product Code MRA
UDI-Device Identifier00888912087414
UDI-Public(01)00888912087414
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
D980003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/20/2010
Device Catalogue Number498-01-048
Device Lot Number957531A
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/29/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/22/2016
Supplement Report FDA Received Date02/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age69 YR
-
-