CORDIS CORPORATION 466FXXXX; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number 466FXXXX
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2011
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review could not be conducted as the sterile lot number was not provided.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
It was reported that an optease vena cava filter was placed in the patient in 2006 and after a while, there was a migration of cava filter in the right ventricle of the heart which was found in 2011.There were no patient injuries.The customer would like to know how long cava filter can maintain its integrity.
 
Manufacturer Narrative
It was reported that an optease vena cava filter was placed in the patient in 2006 and ¿after a while¿, it migrated to the right ventricle of the heart which was found in 2011.There were no patient injuries.The customer would like to know how long cava filter can maintain its integrity.Multiple attempts to gather additional information and contact the customer has been unsuccessful.The product was not returned for analysis.Additionally, as the sterile lot number was not available, device history record review could not be performed.  the optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Retrieval of the optease vena cava filter is indicated up to 23 days post implantation.Following this period of time, it is intended to be permanently implanted.Without images or procedural films for review, the reported filter migration could not be confirmed and the exact cause could not be determined.Giving the limited information available at this time, clinical factor contributing to the migration could not be determined.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration includes mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Given the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466FXXXX
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
p.o. box 025700
miami FL 33152
Manufacturer (Section G)
CORDIS CORPORATION
p.o. box 025700
miami FL 33152
Manufacturer Contact
cecil navajas
14201 nw 60th avenue
miami, FL 33014
MDR Report Key6202214
MDR Text Key63209497
Report Number1016427-2016-00139
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466FXXXX
Device Catalogue Number466FXXXX
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date12/01/2016
Initial Date Manufacturer Received 12/01/2016
Initial Date FDA Received12/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-