• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KANEKA CORPORATION LIXELLE BETA2-MICROGLOBULIN APHERESIS COLUMN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KANEKA CORPORATION LIXELLE BETA2-MICROGLOBULIN APHERESIS COLUMN Back to Search Results
Catalog Number S-35
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fever (1858); Low Blood Pressure/ Hypotension (1914); Tachycardia (2095); Chills (2191)
Date of Event 11/28/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This patient had been tolerated well with lixelle until 26th treatments.The cause of fever, chills, tachycardia is unknown.After administration of antibiotics, these symptoms ceased and there is a possibility that some kind of infection might have occurred.The column size of lixelle was stepwise increased from s-15, s-25 and then to s-35, and this incident occurred when the largest s-35 was used.The possibility of the decrease in blood pressure accompanied by increase in extracorporeal circulation volume cannot be denied.
 
Event or Problem Description
The patient was a male patient on chronic hemodialysis (hd), and had received hd with lixelle, 3 times per week, since (b)(6) 2016.The column size of the lixelle used was increased from s-15 (150ml), s-25 (250ml) and then to s-35 (350ml).On (b)(6) 2016 (in the previous treatment with lixelle s-35, 2 days before the date of this incident) he alleged symptoms at around 20 min.Before the end of the treatment.He was tachycardic, had more chills with fever (around 101.5 f).He refused to go to er and got chest x ray done at that time.He felt better after getting antibiotics (vanco 1g & cefepime 2g since he was allergic to gentamycin.).Because infection was suspected, a blood culture was done (the results was negative as of (b)(6) 2016).On (b)(6) 2016, it was his 28th treatments with lixelle, where s-35 was used, he complained mild chills with fever of 101.2 f towards the end of the treatment.Vanco 1g & cefepime 2g iv were given after the treatment and his symptoms ceased.A blood culture was done again.He has had fluid overload since the previous treatment (700ml positive because he was given considerable amount of normal saline post hd due to hypotension).However, he had remained hypotensive.Sbp checked between 70 and 80mmhg (lying down).O2 sat 94% with oxygen 3l/min via nc.He decided to go to er.He was hospitalized on (b)(6) 2016 and has being hospitalized since then.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIXELLE BETA2-MICROGLOBULIN APHERESIS COLUMN
Common Device Name
BETA2-MICROGLOBULIN APHERESIS COLUMN
Manufacturer (Section D)
KANEKA CORPORATION
2-3-18 nakanoshima, kita-ku
osaka-city, osaka 530-8 288
JA  530-8288
MDR Report Key6203743
Report Number9614654-2016-00024
Device Sequence Number10857187
Product Code PDI
UDI-Device Identifier14993478020218
UDI-Public14993478020218
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H130001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberS-35
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2016
Initial Report FDA Received Date12/25/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age58 YR
-
-