| Model Number |
HGK1608 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
12/05/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint device history records, including collagen coating records, indicated that the graft was processed and inspected according to procedures and no anomaly was found.Specifically, the review of the water permeability testing records of products coated on the same day and under the same conditions as the complaint device indicated values well within product specifications.One retention sample coated on the same period and under the same conditions as the complaint device underwent water permeability testing at 120mmhg as per iso 7198.The test result indicated a value well within product specifications.No conclusion can be drawn.However, all available information and the product testing performed would tend to indicate that the device was not defective.
|
| |
|
Event or Problem Description
|
|
During an aorto bi-femoral intervention performed on (b)(6) 2016 it was reported that a blood leaking occurred through the graft.Hemostatic agent was used and protamine was administrated as antidotum on heparinisation of the patient in order to stop the bleeding.The involved graft remained implanted.No adverse consequence for the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|