• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS HEMAGARD KNITTED; VASCULAR POLYESTER GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS HEMAGARD KNITTED; VASCULAR POLYESTER GRAFT Back to Search Results
Model Number HGK1608
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 12/05/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the complaint device history records, including collagen coating records, indicated that the graft was processed and inspected according to procedures and no anomaly was found.Specifically, the review of the water permeability testing records of products coated on the same day and under the same conditions as the complaint device indicated values well within product specifications.One retention sample coated on the same period and under the same conditions as the complaint device underwent water permeability testing at 120mmhg as per iso 7198.The test result indicated a value well within product specifications.No conclusion can be drawn.However, all available information and the product testing performed would tend to indicate that the device was not defective.
 
Event or Problem Description
During an aorto bi-femoral intervention performed on (b)(6) 2016 it was reported that a blood leaking occurred through the graft.Hemostatic agent was used and protamine was administrated as antidotum on heparinisation of the patient in order to stop the bleeding.The involved graft remained implanted.No adverse consequence for the patient was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD KNITTED
Common Device Name
VASCULAR POLYESTER GRAFT
Manufacturer (Section D)
INTERVASCULAR SAS
zi athelia 1
la ciotat cedex, 13705
FR  13705
MDR Report Key6203821
Report Number1640201-2016-00038
Device Sequence Number19064848
Product Code DSY
UDI-Device Identifier00384401014027
UDI-Public00384401014027
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K964625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/30/2019
Device Model NumberHGK1608
Device Catalogue NumberHGK1608
Device Lot Number14G03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/26/2016
Supplement Report FDA Received Date04/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-