| Model Number |
MAJ-1258 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/08/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject foot switch maj-1258 was returned to olympus for evaluation.Olympus confirmed the subject device, and found there was abnormality of the subject device.There was no considerable damage on the outside of the subject device.A foot switch maker is going to investigate the subject foot switch.There were no further details provided at this time.If significant additional information is received, this report will be supplemented.
|
| |
|
Event or Problem Description
|
|
During the panhysterectomy with the subject foot switch maj-1258 and ues-40, the user facility pressed the cut pedal of subject foot switch to activate the high frequency output, however the high frequency output was not activated.The user facility cycled the power of the subject ues-40 and the user facility pressed the cut pedal of subject foot switch to activate the high frequency output, however the high frequency output was not activated.The user facility pressed the coagulation pedal of subject foot switch to activate the high frequency output, the subject ues-40 normally activated the high frequency output.The user facility replaced the subject ues-40 with the surgical tissue management system (esg-400) and completed the procedure.There was no report of the patient injury other than replacing the device.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report for mfr report #8010047-2016-01609 to provide the device evaluation results.The foot switch maker evaluated the subject foot switch and found that it had no abnormality.The cut pedal and the coagulation pedal of the subject foot switch operated correctly.The subject foot switch was returned to olympus.Olympus evaluated the foot switch again and found that it had no abnormality and operated correctly.There were no further details provided at this time.If significant additional information is received, this report will be supplemented.
|
| |
|
Search Alerts/Recalls
|