| Model Number |
85419 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/13/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
On completion of the investigation a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
During a procedure, the icast stent was deployed successfully.When removing the catheter, the balloon became hung up inside the sheath.When the doctors tried to force the removal, the y hub at the distal end of the catheter broke forcing the doctors to remove the device and the sheath at the same time.
|
| |
|
Additional Manufacturer Narrative
|
|
Engineering investigation: the details state the following: when removing the catheter, the balloon became hung up inside the sheath.It was an 8fr raabe sheath.When the doctors tried to force the removal, the distal end of the catheter broke off.The y hub broke and the doctors had to remove the device and the sheath at the same time.A full review of the catheter lot history records for the devices in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath; ability to deploy the stent at nominal pressure (8atm); ability to withdraw the deflated balloon catheter back through the labeled introducer sheath; ability of the delivery system to withstand 5 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures; balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm); manifold to shaft tensile testing: manifold must not separate at a force less than 3.37lbs.The minimum value recorded from this lot was 7.5lbs.Result: all quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing.The device was removed from the packaging and inspected.Upon initial inspection the catheter hub was not present.Only the broken shaft with the deflated balloon was returned.The sheath used in the case was not returned.The balloon of the catheter appeared to be fully deflated with no signs of failure.The catheter shaft was measured from the tip of the catheter on the distal to the break point of the shaft.The length was 128.5cm.This length is approximately 5cm longer than the usable length specification.This suggests that the catheter was stretched 5cm prior to the shaft breaking.The manifold to shaft tensile data of 29 samples tested had an average break force of 8.49lbs with a minimum value of 7.5lbs.All samples broke in the manifold and were not failures of the adhesive bond.In all cases the shaft necked down and broke in the manifold where the adhesive bond starts.The instructions for use specify the following: "do not use a kinked bronchoscope or endotracheal tube as this may increase the force necessary to deploy the device and may cause a deployment failure or catheter breakage upon removal." conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the lot of stent delivery systems in question.As this device was used in conjunction with an aortic aneurysm graft and a brachial approach, it is possible that the sheath had been kinked or damaged during the procedure.Clinical evaluation: the procedure was a brachial approach to open the internal iliac artery while implanting an abdominal aortic aneurysm graft.After deploying the stent the balloon could not be removed from the sheath.Force was used to remove it and the hub of the catheter broke off.There are several possibilities that could cause a piece of the device to break off including excessive manipulation and force.The instructions for use instruct "do not attempt to withdraw or reposition a balloon catheter within the lumen of the deployed device unless the balloon is completely deflated.".
|
| |
|
Search Alerts/Recalls
|