• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Device Tipped Over (2589); Insufficient Information (3190)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); No Information (3190); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4). It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "vena cava perforation, posterior prongs penetrating walls, projecting beyond the lumen of the inferior vena cava, pain, collapsed lungs". Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. Filter perforation of the vena cava wall is a known risk reported in the published scientific literature. Also, published scientific literature describes that manipulation in the area of filter placement could contribute to changes to the filter configuration and placement thereby potentially initiate perforation of the vena cava wall. There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae. E. G. Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae. In contrast, perforation of adjacent organs is reported with clinical sequelae. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Manufacturer Narrative
Manufacturer ref# (b)(4). Catalog#: unknown but referred to as a cook celect filter. Since catalog# is unknown the 510(k) could be either k061815, k073374, k090140, k112119 or k121057. (b)(4). Investigation is still in progress.
 
Event Description
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2012. " patient outcome: it is alleged that [pt] was injured without further explanation. Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4). Catalog#: unknown but referred to as a cook celect filter. Since catalog# is unknown the 510(k) could be either k061815, k073374, k090140, k112119 or k121057. (b)(4). It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date. Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c. F. R. 803. 56.
 
Event Description
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2012. " patient outcome: it is alleged that [pt] was injured without further explanation. Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4). The event is currently under investigation. A supplemental report will be provided upon conclusion.
 
Event Description
This additional information received on 02/15/2017 as follows: patient was allegedly treated for injuries resulting from motor vehicle accident and claims to have received an implant on (b)(6) 2012. Patient is alleging vena cava perforation, posterior prongs penetrating walls, projecting beyond the lumen of the inferior vena cava, pain, collapsed lungs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA 4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov 4632, IN 47404
8004574500
MDR Report Key6208738
MDR Text Key115720833
Report Number3002808486-2016-01534
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Distributor Facility Aware Date12/06/2016
Event Location No Information
Date Manufacturer Received11/10/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/28/2016 Patient Sequence Number: 1
-
-