Catalog Number IGTCFS-65-JUG-CELECT-PERM
Device Problems Difficult to Remove (1528); Insufficient Information (3190)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); No Information (3190)
Event Date 04/19/2010
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog#: unknown but referred to as a cook celect filter.Since catalog# is unknown the 510(k) could be either k061815, k073374 or k090140.(b)(4).Investigation is still in progress.
 
Event Description
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2009." patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).Catalog#: unknown but referred to as a cook celect filter.Since catalog# is unknown the 510(k) could be either k061815, k073374 or k090140.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2009." patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
Per additional information provided on 02/13/2017: it is alleged that the filter is titled and hook is embedded in the ivc wall.It is reported that the filter is unable to be retrieved.
 
Manufacturer Narrative
(b)(4).According to device history records, based on the lot number provided, there is no evidence to suggest that this device was not manufactured according to specifications or that the filter did not perform as intended, e.G., the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava in the situations described in the ifu.
 
Manufacturer Narrative
(b)(4).Pma 510(k): k073374.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "the filter is tilted and hook is embedded in the ivc wall.It is reported that the filter is unable to be retrieved.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook medical will continue to monitor for similar events.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).Manufacturer reference #(b)(4).Corrected data based on new information received: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "the filter is tilted and hook is embedded in the ivc wall.It is reported that the filter is unable to be retrieved.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook medical will continue to monitor for similar events.Exemption number (b)(4).(b)(4).Manufacturer reference #(b)(4).Corrected data based on new information received: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "the filter is tilted and hook is embedded in the ivc wall.It is reported that the filter is unable to be retrieved.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook medical will continue to monitor for similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer Contact
thomas hessner kirk
400 daniels way
bloomington, IN 47404
8004574500
MDR Report Key6208744
MDR Text Key63407112
Report Number3002808486-2016-01547
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073374
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-JUG-CELECT-PERM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date12/13/2016
Device Age16 MO
Date Manufacturer Received05/23/2017
Date Device Manufactured12/18/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age53 YR
Patient Weight88
-
-