• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012451-12
Device Problems Inflation Problem (1310); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a proximal right coronary artery.A 3.5 x 12 mm xience alpine was deployed in the lesion.A 3.5 x 12 mm nc trek balloon catheter was going to be used for post-dilatation when there was resistance removing the protective sheath.The catheter was advanced to the lesion and when attempting to inflate the balloon, it would not inflate.The catheter was removed and another 3.5 x 12 mm nc trek balloon catheter was used to successfully complete the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Correction: return status changed to not returning.The device was not returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6209427
MDR Text Key63710534
Report Number2024168-2016-09284
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number1012451-12
Device Lot Number61003G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2016
Initial Date FDA Received12/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STENT: 3.5 X 12 XIENCE ALPINE
Patient Age53 YR
Patient Weight94
-
-