• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 55840007560
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Image review: submitted images appear to display mas separated from the bone screw, and two driver tips sheared off at the distal tip.
 
Event or Problem Description
It was reported during surgery, the pedicle screw broke when the surgeon tried to remove it.The product came in contact with patient.Fragments of the broken screw remained in patient.Patient complications are unknown.
 
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Procedure: spondylodesis posterolateralis l5-s1, decompression at l5/s1 cum laminectomy no patient complications were reported as a result of the event.
 
Additional Manufacturer Narrative
Product analysis: visual inspection confirms only the head assembly of the implant returned for analysis, with the (-01) bone screw component missing and not returned for analysis.The ring (-04) and crown (-03) were still assembled in the head.Microscopic examination of the (-04) ring identified ring id feature material displacement, enlarging the id, and the "ears" of the ring bent upward, thus reducing the mechanical retention of the bone screw head.The ring is noted to be fully seated in the retaining groove of the head.The above observations are consistent with mechanical overload as the mechanism of failure.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key6212075
Report Number1030489-2016-03602
Device Sequence Number141695
Product Code NKB
UDI-Device Identifier00643169004375
UDI-Public00643169004375
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number55840007560
Device Lot NumberH5156171
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2017
Initial Date Received by Manufacturer 12/05/2016
Supplement Date Received by ManufacturerNot provided
Not provided
03/02/2017
Initial Report FDA Received Date12/29/2016
Supplement Report FDA Received Date02/20/2017
03/03/2017
09/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
Patient Weight89
-
-