• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. MESA SPINAL SYSTEM; PEDICLE SCREW SPINAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. MESA SPINAL SYSTEM; PEDICLE SCREW SPINAL SYSTEM Back to Search Results
Catalog Number 801-36530M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Information (3190)
Date of Event 11/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A comprehensive investigation was immediately initiated on receipt of the complaint.The product was not returned to manufacturer for evaluation, and a thorough investigation could not be completed as the lot number has not been identified/confirmed in this case.Since the screw remains in the patient, no physical, chemical evaluation could be performed, and the root cause of the reported issue could not be ascertained.It was reported that multiple instruments could not engage and fully lock the screw.It is likely that the lips of outer collet became compromised at some point during the case.Once comprised, it becomes difficult for instruments to fully engage onto the screw head.However, without the actual implant and instruments used in the case no definitive root cause can be determined.While the implant was not returned, a general review of the manufacturing and inspection records revealed no additional information.
 
Event or Problem Description
On 11/30/2016, it was reported to k2m, inc.That a surgery took place in which a mesa screw was left implanted in a partially locked state.Surgery took place (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESA SPINAL SYSTEM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
MDR Report Key6212345
Report Number3004774118-2016-00111
Device Sequence Number19187101
Product Code NKB
UDI-Device Identifier10888857140189
UDI-Public10888857140189
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K133944
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number801-36530M
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2016
Initial Report FDA Received Date12/29/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-