Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Inadequate storage conditions (excessive light and heat), sterilization and/or expired bands could contribute to the properties exhibited during the product evaluation the lot number associated with this complaint was researched to determine the manufacturing date of the bands.We can conclude that the bands were within the acceptable age date range to ensure the bands maintain their elasticity properties.Per the band supplier, if properly stored, the tubing can maintain its specification properties for five years or longer.The tubing should be stored in containers, which are sealed from airflow and light.These bands are stored at our facility in an air tight container to protect against exposure to ultra violet light.We can conclude that the bands were within the acceptable age date range to ensure the bands maintain their elasticity properties.As a precaution, the instructions for use state: "band ligation may not be effective when applied to small varices." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective actions: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|