Catalog Number 1070225-23 |
Device Problems
Difficult to Remove (1528); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/06/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that during device preparation and prior to use, resistance was met while removing the protective sheath and the stent dislodged from the balloon.The device was not used.There was no patient involvement.A different device was used in the procedure without reported issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).Evaluation summary: only the stent was returned for analysis.The reported stent dislodgement was confirmed.The reported difficulty removing the protective sheath could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
|
|
Search Alerts/Recalls
|