Catalog Number 1012535-80
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspections were performed on the returned device.The reported difficult to deploy was unable to be confirmed as the device was returned disassembled and the thumbwheel gear was not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulty appears to be related to circumstances of the procedure as it is likely that the noted kink on the shaft occurred during procedure which contributed to the reported difficulty deploying the stent.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion with chronic total occlusion (cto) located in the superficial femoral artery (sfa) that is heavily calcified.Resistance was not felt during advancement of the 7.0/80 mm/135 cm absolute pro stent delivery system (sds) through lesion and it crossed successfully.The stent was expanded about 1/4 of the length when the thumbwheel of the device became stuck and would not continue to deploy the stent.The handle was disassembled in order to manually work the internal mechanism to deploy the stent successfully.There were no adverse patient effects or clinically significant delay in the procedure reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6215769
Report Number2024168-2016-09392
Device Sequence Number1419211
Product Code NIO
UDI-Device Identifier08717648176081
UDI-Public(01)08717648176081(17)160826(10)6082661
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number1012535-80
Device Lot Number6082661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2016
Initial Report FDA Received Date12/30/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-