Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas alleging a power (battery life) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
|
Manufacturer Narrative
|
The device has been returned and evaluated by product analysis on 02/13/2017 with the following findings: black box shows dates from (b)(6) 2016.Black box shows evidence of unexplained power on reset "por" events and battery alarms following "por" events.No battery cap or cartridge cap returned.All testing performed with test caps.Pump powers with test battery cap.Battery cap is unable to fully tighten.Battery compartment cracks observed from thread area to case seal and below grip pad.Evidence of moisture contamination inside battery compartment.Current draws within specifications.A "battery life" issue was not duplicated during investigation.Leak test shows leak at battery compartment crack.Removed pump case.Evidence of additional moisture inside pump on battery canister.
|
|