Model Number N/A |
Device Problem
Unsealed Device Packaging (1444)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Current information is insufficient to permit a conclusion as to the cause of the event.(b)(4).Corrective action has been initiated for the reported issue.
|
|
Event Description
|
It was reported that when box of bone cement was opened, the inner package of powder was leaking into the outer packaging.There was no patient injury and no delay in a procedure as a result of the event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Evaluation of returned device found the supplier sealing was opened 1/4 of the length.The manufacture sealing was intact.Device history record (dhr) was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to supplier packaging issue.Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|