Manufacturer Narrative
A product investigation was performed for this device.The actual device was not returned to the manufacturer for evaluation.The root cause of the exchange was attributed to the additional trauma of the nail migration into the knee joint.The radiographic and clinical data were reviewed by a sign orthopedic surgeon.Sign fracture care international continues to monitor these events as part of our post market activities.
 
Event Description
We became aware on (b)(6) 2016, that a sign im nail implanted to repair a fracture was replaced due to additional trauma.The nail migrated into the knee joint.The im nail was replaced with a 11mm x 400mm standard nail per the sign technique manual.Surgeon comment: "slippage of previous sign nail into knee joint-- revision of sign nail.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGN IM NAIL
Type of Device
INTRAMEDULLARY FIXATION ROD
Manufacturer (Section D)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer (Section G)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer Contact
robert schmitt
451 hills street
suite b
richland, WA 99354-5502
5093711107
MDR Report Key6216451
MDR Text Key63687510
Report Number3034525-2016-00325
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeTO
PMA/PMN Number
K022632
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age45 YR
Patient Weight91
-
-