MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP Back to Search Results
Device Problems Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue. It was reported that there was a 069 call service alarm. There was no indication that this issue lead to an adverse event. This is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Manufacturer Narrative
Follow-up # 1 date of submission 3/06/2017. Device evaluation: the device has been returned and evaluated by product analysis on 2/13/2017 with the following findings: during testing, the pump was powered on and emitted a cs 069 call service alarm during the rewind step of the investigation. The call service alarm was recorded in the black box data as a cs 087 failure, indicating a language file corruption due to a failed u39 component on the printed circuit board. Unrelated to the initial complaint, it was observed that the battery compartment was cracked, the pump case was cracked between the keypad and the pump seal, and the display screen was dim and discolored. The battery cap and cartridge cap were not returned with the pump and test caps were used to complete the investigation. Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6216709
MDR Text Key63820970
Report Number2531779-2016-35148
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/09/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received12/09/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-