• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM Back to Search Results
Model Number MT22495
Device Problem Wireless Communication Problem (3283)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2016, that on (b)(6) 2016, the patient had a loss of connection.No additional event or patient information is available.No product was provided for evaluation.The reported event could not be confirmed.A root cause cannot be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The receiver was returned for evaluation.A visual inspection was performed and no defects were found.A review of the downloaded data log did not find any errors related to the customer complaint.A pairing test was performed and the test passed.The reported event of a loss of connection was not confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6216733
MDR Text Key63690212
Report Number3004753838-2016-89804
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000156
UDI-Public(01)00386270000156(241)MT22495(10)5219655(17)NA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberMT22495
Device Catalogue NumberSTR-DR-001
Device Lot Number5219655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age85 YR
-
-