• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC, CAMERA HEAD, NON-AC, HD560H; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC, CAMERA HEAD, NON-AC, HD560H; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 72200561S
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2016
Event Type  No Answer Provided  
Event Description
It was reported the camera head non-ac hd560p camera has an extremely poor image.No patient injury or complications were reported.
 
Manufacturer Narrative
Complaint of poor video image was confirmed.Cause of poor image is a film on the lens of camera head.Origin of film on the lens is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC, CAMERA HEAD, NON-AC, HD560H
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon dr.
austin, TX 
5123585706
MDR Report Key6216870
MDR Text Key63846311
Report Number1643264-2016-00318
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200561S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-