Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The evaluation of the photos provided from the customer, show the device catheter and tip.A determination of incorrect orientation cannot be made based off the photos provided.Without return of the complaint device, further investigation cannot be completed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Improper cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is pre-curved and is provided with a pre-curved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of the sphincterotome to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user to "uncoil and straighten sphincterotome" upon removing the device from the packaging.The user is then instructed to "carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements corrective actions: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During preparation for an endoscopic retrograde cholangiopancreatography (ercp), the physician used tri-tome pc triple lumen sphincterotome.The doctor stated that the sphincterotome was not in condition for use because of it's tip deformity.The catheter had no ideal curvature [incorrect cutting wire orientation] to perform the papillotomy.To finalize the procedure, a new product was opened.
|