On (b)(6) 2016, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.There was no indication that the product caused or contributed to an adverse event.
|
Follow-up #1: date of submission 02/21/2017.Device evaluation: the device has been returned and evaluated by product analysis on 01/30/2017 with the following findings: unrelated to the original complaint, the pump powered on to a dim and discolored display.Unrelated to the original complaint, the battery compartment was observed to be cracked.Additionally, the audio bolus button cover was found to be punctured but still responsive to user input.
|