| Catalog Number |
486614110 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/09/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; during surgery (l5/s tlif), the screw of the connector broke when the surgeon striked the screw because it was difficult to fit the connector to the rod.The surgeon used other connector and finished the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection; device history review; complaint history review; risk assessment.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.The most likely cause of the reported event was determined to be the applied overloading while the surgeon hit the connector.
|
| |
|
Event or Problem Description
|
|
It was reported that; during surgery (l5/s tlif), the screw of the connector broke when the surgeon striked the screw because it was difficult to fit the connector to the rod.The surgeon used other connector and finished the surgery.
|
| |
|