Model Number M3536A |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A follow up report will be submitted once the investigation is complete.
|
|
Event Description
|
It was reported to philips that when the user selects the 12 lead command, the device will return to the main screen.The customer is requesting that the device be returned for bench repair.There was no reported patient involvement/adverse patient impact.
|
|
Manufacturer Narrative
|
One processor pca was returned for failure analysis.The reported problem was duplicated in the lab.The device failed 12-leads ecg tests and was unable to analyze 12-leads ecg data after data acquisition.The available information is consistent with a malfunction of the processor pca.The cause was watchdog software-related.The available information from this report does not support that this malfunction represents a systemic, design, or labeling problem.No further investigation or action is warranted.
|
|
Event Description
|
It was reported to philips that when the user selects the 12 lead command, the device will return to the main screen.There was no reported patient involvement/adverse patient impact.
|
|