MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
It was reported to philips that when the user selects the 12 lead command, the device will return to the main screen.The customer is requesting that the device be returned for bench repair.There was no reported patient involvement/adverse patient impact.
 
Manufacturer Narrative
One processor pca was returned for failure analysis.The reported problem was duplicated in the lab.The device failed 12-leads ecg tests and was unable to analyze 12-leads ecg data after data acquisition.The available information is consistent with a malfunction of the processor pca.The cause was watchdog software-related.The available information from this report does not support that this malfunction represents a systemic, design, or labeling problem.No further investigation or action is warranted.
 
Event Description
It was reported to philips that when the user selects the 12 lead command, the device will return to the main screen.There was no reported patient involvement/adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key6236297
MDR Text Key64712622
Report Number1218950-2017-00230
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-