Taper ii, medwatch sent to the fda on 01/10/2017.The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.Based on the serial number provided, the connector type is assumed to be a taper ii.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the reported events of gerd and dysphagia as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Warnings: patients should be advised that the lap-band ap system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
|
|
Taper ii.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.A visual examination was performed on the device, and noted brown discoloration on the lap-band shell, ring, and port tubing.Brown particles were observed on the inner surface of the shell.Scratches were observed on the access port and port holes.Needle marks were noted on the port septum.An air leak test was performed an no leakage was noted.A fill inspection test was performed an no blockage was observed.The port tubing attached to the access port and band tubing had surgical end cuts, consistent with device removal.
|