• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SUPINE TABLE ATTACHMENT (A2) W/PAD; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SUPINE TABLE ATTACHMENT (A2) W/PAD; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Catalog Number 72201725
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 11/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that the surgeon attached the universal distractor and tried to secure it in the table attachment, but the thread would not tighten.The incident resulted in an extended delay in the procedure.
 
Manufacturer Narrative
Examination was not possible, as the device was not returned.The investigation was limited to the information provided.The investigation could not draw any conclusions about the reported event with the clinical details provided.A review of the device history record was not possible due to a lot number not being provided.No further investigation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPINE TABLE ATTACHMENT (A2) W/PAD
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
5123585706
MDR Report Key6237083
MDR Text Key64723895
Report Number3003604053-2017-00006
Device Sequence Number1
Product Code HST
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72201725
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/09/2016
Initial Date FDA Received01/10/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-